ROI® Reliability
Independent testing completed by several laboratories, including
Orthopaedic Research Laboratories, confirms the reliability of the ROI & MC+ implants
as intervertebral body replacement devices.
Mechanical testing
Testing of the ROI and MC+® implants was completed by
independent laboratories in accordance with ASTM F 2077
(Test Methods for Intervertebral Body Fusion Devices).
This testing confirms that the ROI and MC+ provide the
necessary mechanical properties to provide primary
stability and distribution of loads along the spine
segment. These mechanical properties and design of the
implants provide conditions necessary for fusion.
|
ROI Static Axial Compression
Test |
|
Average Maximum Load |
17014 N |
|
Standard Deviation |
944 N |
|
Range |
15685 – 18481 N |
|
ROI Dynamic Compression Test |
|
Frequency |
5 Hz |
|
Load Cycle |
510 – 5100 N |
|
Number of Cycles |
5,000,000 (without cracking or deformation) |
|
ROI Dynamic Torsion Test |
|
Frequency |
5 Hz |
|
Moment Load Cycle |
+/- 2400 N-mm |
|
Number of Cycles |
5,000,000 (without cracking or deformation) |
Independent testing of ROI and
MC+ implants was completed by Orthopaedic Research
Laboratories, CRITT-MDTS and LNE.